DOI 10.1111/ijlh.12150

Медицинская библиотека · International journal of laboratory hematology · 2013

Initial performance evaluation of the UniCel® DxH slide maker/stainer Coulter® cellular analysis system.

Brown W, Keeney M, Hedley B D

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Despite the advancements in instrumentation within hematology laboratories, there is still a need for review of a peripheral blood film (PBF). For a thorough PBF evaluation, it is critical that a well spread and stained film is available.

METHODS

In this study, we evaluated an automatic slide maker/stainer (DxH-SMS, Beckman Coulter) compared with manually prepared blood films on 124 normal and abnormal samples. The primary goal of the study was to determine whether or not the DxH-SMS was able to consistently and reproducibly prepare and stain blood films of exemplary quality, without carryover between specimens. Additionally, repeatability of white blood cell distribution, comparability of morphology to reference methodologies, and grading of acceptance criteria outlined in the CLSI document H20-A2 were assessed.

RESULTS

Carryover was not an issue and repeatability was within expected limits. There was excellent agreement of the 5-part differential between the automated blood films made by the DxH-SMS compared with the manually prepared reference blood film. There was no difference in identification and enumeration of blasts, variant lymphocytes, or nucleated red blood cells (P < 0.05). Red cell morphology showed excellent agreement.

CONCLUSION

Blood films prepared by the DxH-SMS are of excellent quality, reproducible, and compare well with manually prepared slides. Introduction to our laboratory has improved and standardized slide quality.

Для цитирования: Brown W, Keeney M, Hedley B D. Initial performance evaluation of the UniCel® DxH slide maker/stainer Coulter® cellular analysis system.. International journal of laboratory hematology. 2013. DOI: 10.1111/ijlh.12150.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.