Position-Dependent Dysfunction of Deep Brain Stimulation in Tourette Syndrome: Diagnostic Clues.

Wolf Marc E, Majewski Olaf, Müller-Vahl Kirsten R, Blahak Christian, Schulte Dirk-Michael, Krauss Joachim K

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

Detection of defective deep brain stimulation (DBS) contacts/electrodes is sometimes challenging.

CASE REPORT: We report a patient with Tourette syndrome (TS), who presented with abrupt tic increase and mild generalized headache 9 years after DBS implantation. On the suspicion of a hardware defect, a fracture of the DBS electrode and extension lead was ruled out by radiography and standard implantable pulse generator readouts. Further investigation revealed position-dependent modifiable therapeutic impedances, suggesting an impaired contact of the extension lead/adaptor. After replacement normal impedances were recorded, and the patient fully recovered.

DISCUSSION: In DBS dysfunction with inconspicuous hardware check, position-dependent defects might be suspected.

Для цитирования: Wolf Marc E, Majewski Olaf, Müller-Vahl Kirsten R, Blahak Christian, Schulte Dirk-Michael, Krauss Joachim K. Position-Dependent Dysfunction of Deep Brain Stimulation in Tourette Syndrome: Diagnostic Clues.. Tremor and other hyperkinetic movements (New York, N.Y.). 2019. DOI: 10.7916/tohm.v0.713.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.