Pacemaker system failure secondary to air entrapment within the pulse generator pocket. A complication of subclavian venipuncture for lead placement.
Аннотация
Air entrapment within the pulse generator pocket may produce pacemaker system malfunction if the anodal contract plate becomes insulated by the accumulation of air. Unipolar pulse generators are predisposed to this complication. We describe a pacemaker-dependent patient who, early after implantation, experienced pacemaker system failure as a complication of subclavian venipuncture. This patient had an unsuspected pneumothorax that progressed to subcutaneous air entrapment within the pulse generator pocket. Management of this previously unreported complication of subclavian venipuncture is rapid, noninvasive and effective. With the growing use of subclavian venipuncture technique for lead placement one should avoid the predisposing factors that can lead to subcutaneous air entrapment.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.