Chen Chun-Mei, Xia Yun-Cheng, Zhang Xu-Guang, Peng Can-Hui, Liu Fu-You, Peng You-Ming, Sun Lin
Аннотация
AIM: A rapid protocol is necessary to determine the serum concentrations of prednisone.
METHODS: The HP1100 high-performance liquid chromatographic (HPLC) system was employed. The HP Lichrosphere C8 column (250 mm × 4 mm, i.d., 5 μm particle size) was used. The mobile phase was methanol, tetrahydrofuran and water in the ratio 25:25:50. The flow rate was 1.0 ml/min. The sample was monitored by UV absorbance at 240 nm. Acetanilide was used as the internal standard, and methanol was added into the serum for depositing the protein.
RESULTS: The chromatography was effective and was not interfered with by the serum components. Good linearity was observed, within the range of 10-500 μg/L for prednisone, and the detection limit was 5 μg/L. The serum concentrations of prednisone between the nephrotic syndrome (NS) group and the control group were significantly different (P < 0.05), while there was no significant difference between the females and males of the NS group (P > 0.05). The serum ncentration of prednisone in the steroid-resistant group was lower than that in the steroid-sensitive group (P < 0.05).
CONCLUSIONS: HPLC is a practical and reliable method to determine the serum concentration of prednisone with high accuracy, precision, linearity and repeatability.
METHODS: The HP1100 high-performance liquid chromatographic (HPLC) system was employed. The HP Lichrosphere C8 column (250 mm × 4 mm, i.d., 5 μm particle size) was used. The mobile phase was methanol, tetrahydrofuran and water in the ratio 25:25:50. The flow rate was 1.0 ml/min. The sample was monitored by UV absorbance at 240 nm. Acetanilide was used as the internal standard, and methanol was added into the serum for depositing the protein.
RESULTS: The chromatography was effective and was not interfered with by the serum components. Good linearity was observed, within the range of 10-500 μg/L for prednisone, and the detection limit was 5 μg/L. The serum concentrations of prednisone between the nephrotic syndrome (NS) group and the control group were significantly different (P < 0.05), while there was no significant difference between the females and males of the NS group (P > 0.05). The serum ncentration of prednisone in the steroid-resistant group was lower than that in the steroid-sensitive group (P < 0.05).
CONCLUSIONS: HPLC is a practical and reliable method to determine the serum concentration of prednisone with high accuracy, precision, linearity and repeatability.