Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2009

Nuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: optimized protocol for scan time reduction.

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Herzog Bernhard A, Buechel Ronny R, Katz Ruth, Brueckner Michael, Husmann Lars, Burger Irene A, Pazhenkottil Aju P, Valenta Ines, Gaemperli Oliver, Treyer Valerie, Kaufmann Philipp A

Аннотация

UNLABELLED: We aimed at establishing the optimal scan time for nuclear myocardial perfusion imaging (MPI) on an ultrafast cardiac gamma-camera using a novel cadmium-zinc-telluride (CZT) solid-state detector technology.

METHODS: Twenty patients (17 male; BMI range, 21.7-35.5 kg/m(2)) underwent 1-d (99m)Tc-tetrofosmin adenosine stress and rest MPI protocols, each with a 15-min acquisition on a standard dual-detector SPECT camera. All scans were immediately repeated on an ultrafast CZT camera over a 6-min acquisition time and reconstructed from list-mode raw data to obtain scan durations of 1 min, 2 min, etc., up to a maximum of 6 min. For each of the scan durations, the segmental tracer uptake value (percentage of maximum myocardial uptake) from the CZT camera was compared by intraclass correlation with standard SPECT camera data using a 20-segment model, and clinical agreement was assessed per coronary territory. Scan durations above which no further relevant improvement in uptake correlation was found were defined as minimal required scan times, for which Bland-Altman limits of agreement were calculated.

RESULTS: Minimal required scan times were 3 min for low dose (r = 0.81; P < 0.001; Bland-Altman, -11.4% to 12.2%) and 2 min for high dose (r = 0.80; P < 0.001; Bland-Altman, -7.6% to 12.9%), yielding a clinical agreement of 95% and 97%, respectively.

CONCLUSION: We have established the minimal scan time for a CZT solid-state detector system, which allows 1-d stress/rest MPI with a substantially reduced acquisition time resulting in excellent agreement with regard to uptake and clinical findings, compared with MPI from a standard dual-head SPECT gamma-camera.