Miniaturized Photo-Ionization Fourier Deconvolution Ion Mobility Spectrometer for the Detection of Volatile Organic Compounds.

Yang Binwang, Yu Jianna, Liu Wen, Jing Guoxing, Li Wenshan, Liu Wenjie

СтатьяПеревод в очереди

Аннотация

Because of its simplicity, reliability, and sensitivity, the drift tube ion mobility spectrometer (IMS) has been recognized as the equipment of choice for the on-site monitoring and identification of volatile organic compounds (VOCs). However, the performance of handheld IMS is often limited by the size, weight, and drift voltage, which heavily determine the sensitivity and resolving power that is crucial for the detection and identification of VOCs. In this work, we present a low-cost, miniaturized drift tube ion mobility spectrometer incorporated with a miniaturized UV ionization lamp and a relatively low drift voltage. The sensitivity and resolving power are boosted with the implementation of Fourier deconvolution multiplexing compared to the conventional signal averaging data acquisition method. The drift tube provides a high resolving power of up to 52 at a drift length of 41 mm, 10 mm ID dimensions, and a drift voltage of 1.57 kV. Acetone, benzene, dimethyl methyl phosphonate, methyl salicylate, and acetic acid were evaluated in the developed spectrometer and showed satisfactory performance.

Для цитирования: Yang Binwang, Yu Jianna, Liu Wen, Jing Guoxing, Li Wenshan, Liu Wenjie. Miniaturized Photo-Ionization Fourier Deconvolution Ion Mobility Spectrometer for the Detection of Volatile Organic Compounds.. Sensors (Basel, Switzerland). 2022. DOI: 10.3390/s22155468.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.