Inhaled nitric oxide (iNO) delivery with high-frequency jet ventilation (HFJV).
Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Platt David R, Swanton Doug, Blackney David
Аннотация
OBJECTIVE: To determine if nitric oxide (NO) therapy can be reliably administered during high-frequency jet ventilation (HFJV) using the INOvent delivery system.
STUDY DESIGN: NO concentrations were measured just proximal to the endotracheal (ET) tube and at the distal tip of the ET tube during a bench evaluation. Measurements were taken over a wide range of airway pressure settings and NO concentrations with both high- and low- resistance lung models. Percent changes in set versus proximal and proximal versus distal iNO concentrations were tabulated.
RESULTS: Differences between proximal and distal NO concentrations were 10% or less. In the therapeutic range of up to 20 p.p.m., differences in concentration were 1 p.p.m. or less. There was no consistent effect on NO concentration when airway resistance was increased by 500%.
CONCLUSION: Nitric oxide therapy can be reliably administered during HFJV with the INOvent delivery system when NO is injected exclusively via the HFJV circuit.
STUDY DESIGN: NO concentrations were measured just proximal to the endotracheal (ET) tube and at the distal tip of the ET tube during a bench evaluation. Measurements were taken over a wide range of airway pressure settings and NO concentrations with both high- and low- resistance lung models. Percent changes in set versus proximal and proximal versus distal iNO concentrations were tabulated.
RESULTS: Differences between proximal and distal NO concentrations were 10% or less. In the therapeutic range of up to 20 p.p.m., differences in concentration were 1 p.p.m. or less. There was no consistent effect on NO concentration when airway resistance was increased by 500%.
CONCLUSION: Nitric oxide therapy can be reliably administered during HFJV with the INOvent delivery system when NO is injected exclusively via the HFJV circuit.