MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/01.pcc.0000059429.95034.73

Медицинская библиотека · Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies · 2003

Correction of compliance and resistance altered by endotracheal tube leaks.

Nikischin Werner, Lange Malte

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

Measurements of lung compliance and resistance are influenced by endotracheal tube leaks. To keep compliance and resistance reliable, we developed an algorithm to correct inspiratory and expiratory volume and flow mathematically.

DESIGN

Prospective, clinical study.

SETTING

University research laboratory and neonatal intensive care unit.

MODEL: A ventilated lung model with a linear pressure-volume relationship and with adjustment of an increasing endotracheal tube leak was investigated.

PATIENTS: A total of 21 ventilated premature neonates (median weight, 1220 g; range, 640-2160 g; median leak, 32%; range, 24-56%) were studied.

MEASUREMENTS AND MAIN

RESULTS

Compliance and resistance were calculated from the recordings of flow, volume, and airway pressure over time employing linear regression of the equation of motion to obtain compliance and resistance. Compliance and resistance altered by leaks were corrected and compared with measurements without leak. Compliance and resistance of the lung model could be corrected up to an endotracheal tube leak size of 86%. Compliance and resistance without leak and after leak correction did not differ significantly for all infants using the linear regression method (p >.05). For the correction of compliance in 15 and for the correction of resistance in 12 of the 21 infants, the coefficients of variation of ten measured breaths without leak were greater or equal to the differences of the values of compliance and resistance between conditions of no leak and corrected leak, respectively.

CONCLUSION

Pulmonary compliance and resistance can be reliably corrected even in the presence of a substantial endotracheal tube leak, which makes pulmonary function tests more reliable.

Для цитирования: Nikischin Werner, Lange Malte. Correction of compliance and resistance altered by endotracheal tube leaks.. Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies. 2003. DOI: 10.1097/01.pcc.0000059429.95034.73.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.