DOI 10.3109/00365513.2014.939995
Laboratory and clinical evaluation of on-site urine drug testing.
Аннотация
AIM: Products for on-site urine drug testing offer the possibility to perform screening for drugs of abuse directly at the point-of-care. This is a well-established routine in emergency and dependency clinics but further evaluation of performance is needed due to inherent limitations with the available products.
METHODS
Urine drug testing by an on-site product was compared with routine laboratory methods. First, on-site testing was performed at the laboratory in addition to the routine method. Second, the on-site testing was performed at a dependency clinic and urine samples were subsequently sent to the laboratory for additional analytical investigation.
RESULTS
The on-site testing products did not perform with assigned cut-off levels. The subjective reading between the presence of a spot (i.e. negative test result) being present or no spot (positive result) was difficult in 3.2% of the cases, and occurred for all parameters. The tests performed more accurately in drug negative samples (specificity 96%) but less accurately for detecting positives (sensitivity 79%). Of all incorrect results by the on-site test the proportion of false negatives was 42%. The overall agreement between on-site and laboratory testing was 95% in the laboratory study and 98% in the clinical study.
CONCLUSION
Although a high degree of agreement was observed between on-site and routine laboratory urine drug testing, the performance of on-site testing was not acceptable due to significant number of false negative results. The limited sensitivity of on-site testing compared to laboratory testing reduces the applicability of these tests.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.