Determination of normal hepatic elasticity by using real-time shear-wave elastography.
Аннотация
PURPOSE: To determine normal reference values of liver elasticity and measurement reliability by using real-time shear-wave elastography (SWE) in patients with a range of ages and body mass index (BMI) measurements, with presence or absence of hepatic steatosis.
MATERIALS AND
METHODS
The institutional review board approved this study, and informed consent was waived because of the retrospective nature of the study. Two hundred thirty-eight patients who underwent SWE and ultrasonography-guided liver biopsies on the same day were identified retrospectively. The median kilopascal value of three consecutive measurements was used as a representative value for each subject. One hundred ninety-six patients who were potential donors for living-donor liver transplantation and had biopsy-proven normal (123 nonsteatotic and 73 steatotic) livers as the only histologic abnormality were included in the study. Reference ranges of normal hepatic elasticity were calculated by using lower and upper limits at the 2.5 and 97.5 percentiles. With the upper value of the reference range as a cutoff value, the sensitivity and specificity for the diagnosis of hepatic fibrosis were calculated. Measurement reliability was evaluated by using the intraclass correlation coefficient (ICC). To investigate the effects of potential confounding factors (age, hepatic steatosis, and BMI) on liver elasticity, the Pearson correlation test and the Student t test were performed.
RESULTS
The reference range of normal hepatic elasticity was 2.6-6.2 kPa. With 6.2 kPa as a cutoff value, the sensitivity and specificity for the diagnosis of hepatic fibrosis were 91% (20 of 22 subjects) and 95.9% (188 of 196 subjects), respectively. The overall ICC for the elasticity measurements was 0.924. The potential confounding factors that we considered had negligible effects on the elasticity values.
CONCLUSION
Hepatic elasticity values measured with SWE in histologically proven normal livers ranged from 2.6 to 6.2 kPa, with high measurement reliability. The effect of the potential confounding factors on liver elasticity was negligible.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.