The validity and reliability of an open source biosensing board to quantify heart rate variability.

Burma Joel S, Lapointe Andrew P, Soroush Ateyeh, Oni Ibukunoluwa K, Smirl Jonathan D, Dunn Jeff F

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Heart rate variability (HRV) is a popular tool to quantify autonomic function. However, this typically requires an expensive 3-12 lead electrocardiogram (ECG) and BioAmp system. This investigation sought to determine the validity and reliability of an OpenBCI cyton biosensing board (open source) for accurately quantifying HRV.

NEW METHOD: A cyton board with a 3-lead ECG was employed to acquire heart rate waveform data, which was processed to obtain HRV within both time- and frequency-domains. The concurrent validity was compared to a simultaneous recording from an industry-standard 3-lead ECG (ADInstruments) (n = 15). The reliability of the cyton board was compared between three days within a 7-day timespan (n = 10). Upright quiet-stance short-term HRV metrics were quantified in time- and frequency-domains.

RESULTS

The two devices displayed excellent limits of agreements (all log mean differences ±0.4) and very high between-device variable associations (all r 2 > 0.98). Between the three time points in the same subjects, no differences were noted within time- (all p > 0.71) or frequency-domains (all p > 0.88) across testing points. Finally, all HRV metrics exhibited excellent levels of reliability through high Cronbach's Alpha (all ≥0.916) and intraclass correlation coefficients (all ≥0.930); and small standard error of the measurement (all ≤0.7) and typical error of the measurement (all ≤0.1) metrics.

COMPARISON WITH EXISTING

METHODS

The cyton board with 3-lead ECG was compared with an industry-standard ADInstruments ECG during HRV assessments. There were no significant differences between devices with respect to time- and frequency-domains. The cyton board displayed high-levels of between-day reliability and provided values harmonious to previous ECG literature highlighting the applicability for longitudinal studies.

CONCLUSION

With proper background knowledge regarding ECG principles and a small degree of set-up complexity, an open source cyton board can be created and employed to perform multimodal HRV assessments at a fraction of the cost (~4%) of an industry-standard ECG setup.

Для цитирования: Burma Joel S, Lapointe Andrew P, Soroush Ateyeh, Oni Ibukunoluwa K, Smirl Jonathan D, Dunn Jeff F. The validity and reliability of an open source biosensing board to quantify heart rate variability.. Heliyon. 2021. DOI: 10.1016/j.heliyon.2021.e07148.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.