Impact of magnetic fields on calculated AAPM TG-43 parameters for 192Ir and 60Co HDR brachytherapy sources: A Monte Carlo study.

Ranjbar Sahar, Gholami Somayeh, Longo Francesco, Sharp Ryan, Meigooni Ali S

СтатьяПеревод в очереди

Аннотация

PURPOSE: The aim of this work is to investigate the influence of an external magnetic field (MF) on The American Association of Physicists in Medicine (AAPM) No. 43 Report (TG-43) parameters for 192Ir and 60Co high dose rate (HDR) brachytherapy sources using Monte Carlo (MC) simulation methods.

MATERIALS AND

METHODS

We used the Geant4 toolkit (version 10.1. p01) to simulate the geometry of 192Ir and 60Co brachytherapy sources. AAPM TG-43 parameters (the radial dose function, g(r), and the anisotropy function, F (r, θ)) of both 192Ir and 60Co sources were calculated in the presence of a magnetic field with strengths of 1.5T, 3T, and 7T in the X, Y, and Z directions in a voxelized water phantom.

RESULTS

For the 192Ir source, the calculated values g(r) and F (r, θ) remained nearly unaffected by the magnetic field for all investigated strengths. For the 60Co source, the differences for the g(r) and F (r,θ) under the 1.5T, 3T, and 7T magnetic field strengths along the direction parallel with the MF were found to be an increase of up to 5%, 15%, and 33%, respectively. However, for the directions perpendicular with the magnetic field, there was a decrease of up to 3%, 6% and 15% under 1.5T, 3T and 7T strengths, respectively.

CONCLUSION

Our results highlight the necessity of a Monte Carlo-based treatment planning system (TPS) if cobalt HDR treatments are performed under a magnetic field, especially for strengths greater than 1.5T.

Для цитирования: Ranjbar Sahar, Gholami Somayeh, Longo Francesco, Sharp Ryan, Meigooni Ali S. Impact of magnetic fields on calculated AAPM TG-43 parameters for 192Ir and 60Co HDR brachytherapy sources: A Monte Carlo study.. Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine. 2020. DOI: 10.1016/j.apradiso.2020.109088.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.