Clinical Efficacy of Bilateral Deep Brain Stimulation Does Not Change After Implantable Pulse Generator Replacement but the Impedances Do: A Prospective Study.

Niemann Marcel, Schneider Gerd-Helge, Kühn Andrea, Vajkoczy Peter, Faust Katharina

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Deep brain stimulation (DBS) is an approved therapy option for movement disorders such as Parkinson's disease (PD), essential Tremor (ET), and dystonia. While current research focuses on rechargeable implantable pulse generators (IPGs), little is known about changes of the motor functions after IPG replacement and the consequences of additionally implanted hardware.

OBJECTIVE

To assess changes of the motor functions, the therapy impedances, and the total electric energy delivered (TEED) after elective IPG replacement.

METHODS

We prospectively acquired the data of 47 patients with PD, ET, and dystonia treated with bilateral DBS. Motor functions were rated prior to and after surgery using the revised Unified Parkinson's Disease Rating Scale part III (MDS-UPDRS-III), the Fahn-Tolosa-Marin Tremor-Rating-Scale (FTM-TRS), and the Unified Dystonia Rating Scale (UDRS). Furthermore, the therapy impedances and TEED were assessed at the aforementioned times.

RESULTS

While preoperative motor scores were 48.32 ± 17.16 in PD, 39.71 ± 12.28 in ET, and 18.48 ± 16.30 in dystonia patients, postoperative scores were 47.84 ± 24.33, 32.86 ± 15.82, and 15.02 ± 15.17, respectively. Only in dystonia patients, motor scores significantly differed. Perioperative therapy impedance changes were 142.66 ± 105.35 Ω (Kinetra® to Activa® PC), -68.75 ± 43.05 Ω (Activa® PC to Activa® PC), and - 51.38 ± 38.75 Ω (Activa® PC to Activa® RC). Perioperative TEED changes were - 37.15 ± 38.87 μJ, 2.03 ± 35.91 μJ, and 12.39 ± 6.31 μJ in that first, second, and third group, respectively. Both the therapy impedances and TEED significantly differed between groups.

CONCLUSION

Although there were no statistically significant changes in the motor functions of all patients after elective IPG replacement, the therapy impedances were significantly higher and TEED was significantly lower after IPG replacement with concurrent Pocket Adapter implantation.

Для цитирования: Niemann Marcel, Schneider Gerd-Helge, Kühn Andrea, Vajkoczy Peter, Faust Katharina. Clinical Efficacy of Bilateral Deep Brain Stimulation Does Not Change After Implantable Pulse Generator Replacement but the Impedances Do: A Prospective Study.. Neuromodulation : journal of the International Neuromodulation Society. 2019. DOI: 10.1111/ner.13022.