MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/jat/5.1.14

Медицинская библиотека · Journal of analytical toxicology · 1981

Evaluation of the TRI 'dipstick' test for the detection of drugs of abuse in urine.

Jukofsky D, Kramer A, Mulé S J

СтатьяПеревод в очереди

Аннотация

An evaluation of Technology Resources Inc. (TRI) Amphetamine, Barbiturate, Narcotic (G) and Narcotic (S) "Dipsticks" for drugs of abuse in urine was made. The results obtained by six individuals reading the "Dipstick" papers was compared with the analysis of the same urine samples, by a combination of TLC, EMIT, RIA and GLC. The data obtained with "Dipstick" papers, regardless of the drug tested, were clearly unreliable (high percentage of false negatives, low percentage of true positives) and the assay was unsuitable as a technique for screening urines for drugs of abuse.

Для цитирования: Jukofsky D, Kramer A, Mulé S J. Evaluation of the TRI 'dipstick' test for the detection of drugs of abuse in urine.. Journal of analytical toxicology. 1981. DOI: 10.1093/jat/5.1.14.