Wei sheng yan jiu = Journal of hygiene research · 2021
[Determination of domoic acid in shellfish by liquid chromatography- tandem mass spectrometry coupled with dispersive solid phase extraction].
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Fang Li, Wu Fan, Qiu Fengmei, Yu Xinwei, Zhu Yanhua
Аннотация
OBJECTIVE: A method utilizing liquid chromatography-tandem mass spectrometry(LC-MS/MS) coupled with dispersive solid phase extraction for quantitative analysis of domoic acid in four kinds of shellfish was established.
METHODS: The sample of 0. 1 g was extracted with 25% methanol aqueous solution, the extract was purified by dispersive solid phase extraction with 50 mg HLB and 5 mg GCB, and then filtered through a PTFE membrane. The analytes were separated on a C_(18) column(100 mm×2. 1 mm, 1. 9 μm), and detected in selected reaction monitoring(SRM) mode via positive electrospray ionization. The matrix matching and external standard method was used for quantitation.
RESULTS: Domoic acid showed good linearity in the concentration range between 1. 0 ng/mL and 50. 0 ng/mL with correlation coefficients higher than 0. 9994. The detection limits of domoic acid in shellfish was 5 μg/kg. The inter-and intra-day recoveries were 91. 6%-109. 2% and 90. 9%-109. 3%, respectively. The inter-and intra-day ralitive standard deviations(RSDs) were lower than 8. 2% at spiked concentrations of 20, 50 and 100 μg/kg.
CONCLUSION: The method is accurate, fast, easy to operate, which can satisfy the requirements of public health emergency testing or routine testing.
METHODS: The sample of 0. 1 g was extracted with 25% methanol aqueous solution, the extract was purified by dispersive solid phase extraction with 50 mg HLB and 5 mg GCB, and then filtered through a PTFE membrane. The analytes were separated on a C_(18) column(100 mm×2. 1 mm, 1. 9 μm), and detected in selected reaction monitoring(SRM) mode via positive electrospray ionization. The matrix matching and external standard method was used for quantitation.
RESULTS: Domoic acid showed good linearity in the concentration range between 1. 0 ng/mL and 50. 0 ng/mL with correlation coefficients higher than 0. 9994. The detection limits of domoic acid in shellfish was 5 μg/kg. The inter-and intra-day recoveries were 91. 6%-109. 2% and 90. 9%-109. 3%, respectively. The inter-and intra-day ralitive standard deviations(RSDs) were lower than 8. 2% at spiked concentrations of 20, 50 and 100 μg/kg.