European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2017
Insertion forces and intracochlear trauma in temporal bone specimens implanted with a straight atraumatic electrode array.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Mirsalehi Marjan, Rau Thomas S, Harbach Lenka, Hügl Silke, Mohebbi Saleh, Lenarz Thomas, Majdani Omid
Аннотация
The aim of the study was to evaluate insertion forces during manual insertion of a straight atraumatic electrode in human temporal bones, and post-implantation histologic evaluation of the samples to determine whether violation of intracochlear structures is related to insertion forces. In order to minimize intracochlear trauma and preserve residual hearing during cochlear implantation, knowledge of the insertion forces is necessary. Ten fresh frozen human temporal bones were prepared with canal wall down mastoidectomy. All samples were mounted on a one-axis force sensor. Insertion of a 16-mm straight atraumatic electrode was performed from different angles to induce "traumatic" insertion. Histologic evaluation was performed in order to evaluate intracochlear trauma. In 4 of 10 samples, dislocation of the electrode into scala vestibuli was observed. The mean insertion force for all 10 procedures was 0.003 ± 0.005 N. Insertion forces measured around the site of dislocation to scala vestibuli in 3 of 4 samples were significantly higher than insertion forces at the same location of the cochleae measured in samples without trauma (p < 0.04). Mean force during the whole insertion process of the straight atraumatic electrode is lower than reported by other studies using longer electrodes. Based on our study, insertion forces leading to basilar membrane trauma may be lower than the previously reported direct rupture forces.