Comparison of Mechanical Properties of PEEK and PMMA: An In Vitro Study.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Shrivastava Spardha P, Dable Rajani, Raj A P Nirmal, Mutneja Puneet, Srivastava Sparsh B, Haque Merazul
Аннотация
AIM: Evaluation and comparison of flexural strength and hardness of heat-cured polymethylmethacrylate (PMMA) denture base with polyetheretherketone (PEEK).
MATERIAL AND METHODS: A total of 80 samples of PEEK and PMMA were prepared and divided into two groups: group A and group B, of 40 each. Both the groups were further divided into subgroups consisting of 20 samples of PMMA and PEEK. The group A specimens were tested for flexural strength by universal testing machine (UTM), and group B samples were subjected to hardness test using Vickers microhardness tester. The values obtained were subjected to statistical analysis. The readings were subjected to an unpaired t-test for comparison of flexural strength and hardness of PEEK and PMMA.
RESULT: The flexural strength of PEEK was 183 MPa, while that of PMMA was 84 MPa. The hardness of PEEK and PMMA was 24 VHN and 19.4 VHN, respectively.
CONCLUSION: It was observed that PEEK can be a potential denture base material with superior properties as compared to PMMA. Further researches are needed to be carried out.
CLINICAL SIGNIFICANCE: PEEK as a denture base exhibits adequate flexural strength prolonging its clinical longevity and overcoming the most common reason for the failure of PMMA denture base that results in fracture.
MATERIAL AND METHODS: A total of 80 samples of PEEK and PMMA were prepared and divided into two groups: group A and group B, of 40 each. Both the groups were further divided into subgroups consisting of 20 samples of PMMA and PEEK. The group A specimens were tested for flexural strength by universal testing machine (UTM), and group B samples were subjected to hardness test using Vickers microhardness tester. The values obtained were subjected to statistical analysis. The readings were subjected to an unpaired t-test for comparison of flexural strength and hardness of PEEK and PMMA.
RESULT: The flexural strength of PEEK was 183 MPa, while that of PMMA was 84 MPa. The hardness of PEEK and PMMA was 24 VHN and 19.4 VHN, respectively.
CONCLUSION: It was observed that PEEK can be a potential denture base material with superior properties as compared to PMMA. Further researches are needed to be carried out.
CLINICAL SIGNIFICANCE: PEEK as a denture base exhibits adequate flexural strength prolonging its clinical longevity and overcoming the most common reason for the failure of PMMA denture base that results in fracture.