A Novel, Low-Cost, Wide-Field, Infant Retinal Camera, "Neo": Technical and Safety Report for the Use on Premature Infants.

Vinekar Anand, Rao Shyam Vasudeva, Murthy Seema, Jayadev Chaitra, Dogra Mangat R, Verma Anshuman, Shetty Bhujang

СтатьяПеревод в очереди

Аннотация

PURPOSE: To report the technical aspects, systemic, and ocular safety of a novel, low-cost, wide-field, infant retinal camera for use on premature infants.

METHODS

The device, the "3nethra Neo" (Neo) is a 120° portable, contact, wide-field, unibody camera, with a CMOS sensor (2040 × 2040 resolution) and a warm light-emitting diode (LED) illumination source. The Neo was used to image 140 awake, preterm infants between postmenstrual age (PMA) of 28 to 37 weeks, undergoing retinopathy of prematurity (ROP) screening. Baseline, 'during procedure', at 5 minutes, and for 60 minutes postprocedure, readings of oxygen saturation and heart rate were recorded. The device design, optics, illumination, and software specifications were compared with the RetCam 3.

RESULTS

Study defined bradycardia (9 infants, 6.4%), tachycardia (3 infants, 2.1%), and hypoxia (2 infants, 1.4%) were observed but there were no clinically significant systemic changes that required intervention during or following any of the study time intervals. There was a transient increase in heart rate by 9.68 (7.53-11.83; P < 0.0001) and marginal decrease in oxygen saturation (-1.94 [-1.60 to -2.28], P < 0.0001), which started to return to baseline 5 minutes after the procedure. Transient redness was seen in two eyes (0.7%) of two infants. No other ocular adverse effects were observed.

CONCLUSIONS

The Neo is easy to use in preterm infants and being compact was readily portable. There were no significant ocular or systemic adverse effects, potentially allowing it to be a viable low-cost device for ROP screening in low resource settings.

TRANSLATIONAL RELEVANCE: The camera provides a safe and affordable alternative to image the retina of infants by using novel illumination and lens mechanics and has the potential of worldwide acceptance.

Для цитирования: Vinekar Anand, Rao Shyam Vasudeva, Murthy Seema, Jayadev Chaitra, Dogra Mangat R, Verma Anshuman, Shetty Bhujang. A Novel, Low-Cost, Wide-Field, Infant Retinal Camera, "Neo": Technical and Safety Report for the Use on Premature Infants.. Translational vision science & technology. 2019. DOI: 10.1167/tvst.8.2.2.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.