Comparison of 3D-printed titanium-alloy, standard titanium-alloy, and PEEK interbody spacers in an ovine model.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Van Horn Margaret R, Beard Roland, Wang Wenhai, Cunningham Bryan W, Mullinix Kenneth P, Allall May, Bucklen Brandon S

Аннотация

BACKGROUND CONTEXT: Osseointegration is a pivotal process in achieving a rigid fusion and ultimately a successful clinical outcome following interbody fusion surgery. Advancements in 3D printing technology permit commonly used titanium interbody spacers to be designed with unique architectures, such as a highly interconnected and specific porous structure that mimics the architecture of trabecular bone. Interbody implants with a microscale surface roughness and biomimetic porosity may improve bony ongrowth and ingrowth compared to traditional materials.

PURPOSE: The purpose of this study was to compare the osseointegration of lumbar interbody fusion devices composed of surgical-grade polyetheretherketone (PEEK), titanium-alloy (TAV), and 3D-printed porous, biomimetic TAV (3DP) using an in vivo ovine model.

STUDY DESIGN: In Vivo Preclinical Animal Study METHODS: Eighteen sheep underwent two-level lateral lumbar interbody fusion randomized with either 3DP, PEEK, or TAV interbody spacers (n=6 levels for each spacer per time point). Postoperative time points were 6 and 12 weeks. Microcomputed tomography and histomorphometry were used to quantify bone volume (BV) within the spacers (ingrowth) and the surface bone apposition ratio (BAR) (ongrowth), respectively.

RESULTS: The 3DP-treatment group demonstrated significantly higher BV than the PEEK and TAV groups at 6 weeks (77.3±44.1 mm3, 116.9±43.0 mm3, and 108.7±15.2 mm3, respectively) (p<.05). At 12 weeks, there were no BV differences between groups (p>.05). BV increased in all groups from the 6- to 12-week time points (p<.05). At both time points, the 3DP-treated group (6w: 23.6±10.9%; 12w: 36.5±10.9%) had significantly greater BAR than the PEEK (6w: 8.6±2.1%; 12w: 14.0±5.0%) and TAV (6w: 6.0±5.7%; 12w: 4.1±3.3%) groups (p<.05).

CONCLUSIONS: 3DP interbody spacers facilitated greater total bony ingrowth at 6 weeks, and greater bony ongrowth postoperatively at both 6 and 12 weeks, in comparison to solid PEEK and TAV implants.

CLINICAL SIGNIFICANCE: Based on these findings, the 3DP spacers may be a reasonable alternative to traditional PEEK and TAV spacers in various clinical applications of interbody fusion.