MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/spv.0000000000001105

Медицинская библиотека · Female pelvic medicine & reconstructive surgery · 2022

Six-Month Outcomes of Reimplantation of a Coin-Sized Tibial Nerve Stimulator for the Treatment of Overactive Bladder Syndrome With Urgency Urinary Incontinence.

Kaaki Bilal, English Sharon, Gilling Peter, Meffan Patrick, Lucente Vincent, MacDiarmid Scott, Clark Matthew, Sen Subhro K

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The eCoin (Valencia Technologies Corporation, Valencia, CA) is a battery-powered, nickel-sized and shaped neuromodulation device for the treatment of overactive bladder, and it is implanted in the lower leg in a short office or outpatient procedure under local anesthesia. A follow-on trial was conducted to evaluate the feasibility, safety, and efficacy of eCoin reimplantation.

METHODS

This was a prospective, single-arm, open-label study, including 23 participants with refractory urgency urinary incontinence (UUI) who were previously participants in the eCoin clinical feasibility trial. This follow-on study was conducted at 7 sites in the United States and New Zealand. Participants were reimplanted with a new eCoin device and activated after 4 weeks. Bladder diary data and validated quality-of-life instruments, collected at 12 weeks and 24 weeks postactivation, were compared with baseline.

RESULTS

Participants of the study were considered responders if they reported a ≥ 50% reduction from baseline in episodes of UUI on a 3-day voiding diary. At 12 weeks of treatment, 74% (95% confidence interval [CI], 52%-90%) of participants were considered responders. At 24 weeks of treatment, 82% (95% CI, 60%-95%) of participants were considered responders with 36% (95% CI, 20%-57%) of participants achieving complete continence. There were no device-related serious adverse events reported during the study.

CONCLUSIONS

The reimplantation of eCoin was both safe and effective in treating UUI associated with overactive bladder syndrome. The demonstrated significant reduction or resolution of symptoms with no serious safety concern suggests that eCoin is a convenient and maintainable therapeutic device.

Для цитирования: Kaaki Bilal, English Sharon, Gilling Peter, Meffan Patrick, Lucente Vincent, MacDiarmid Scott, Clark Matthew, Sen Subhro K. Six-Month Outcomes of Reimplantation of a Coin-Sized Tibial Nerve Stimulator for the Treatment of Overactive Bladder Syndrome With Urgency Urinary Incontinence.. Female pelvic medicine & reconstructive surgery. 2022. DOI: 10.1097/spv.0000000000001105.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.