Persistent hemolysis after transcatheter occlusion of a patent ductus arteriosus: surgical ligation of the duct over the occlusion device.
Клинический случайСтатьяПеревод в очереди
Аннотация
A 14-month-old child underwent routine transcatheter closure of her patent ductus arteriosus (PDA) using the Rashkind double-umbrella device. She developed severe persistent hemolysis in association with residual ductal flow. The problem was managed by the previously unreported approach of standard surgical ligation of the duct via a left thoracotomy, leaving the occlusion device in situ.
Для цитирования: Chisholm J C, Salmon A P, Keeton B R, Webber S A, Monro J L. Persistent hemolysis after transcatheter occlusion of a patent ductus arteriosus: surgical ligation of the duct over the occlusion device.. Pediatric cardiology. 1995. DOI: 10.1007/bf00794194.