Safety and feasibility of a midseptal implantation technique of a leadless pacemaker.

Hai Jo-Jo, Fang Jonathan, Tam Chor-Cheung, Wong Chun-Ka, Un Ka-Chun, Siu Chung-Wah, Lau Chu-Pak, Tse Hung-Fat

СтатьяНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The major risk of implanting a leadless pacemaker at the right ventricular (RV) apex is cardiac perforation.

OBJECTIVE

The purpose of this study was to describe and prospectively evaluate the safety and feasibility of a technique for midseptal implantation of the Micra leadless pacemaker.

METHODS

We positioned the device at the center of the cardiac silhouette in the right anterior oblique (RAO) view, toward the left in the left anterior oblique (LAO) view, and away from the sternum in the left lateral view.

RESULTS

Among the 51 patients (mean age 81.3 ± 9.3 years; 47% men) included in the study, 29 (57%) were >80 years old, 7 (14%) had body mass index <20 kg/m2, 48 (94%) had renal dysfunction, and 33 (65%) had valvular heart disease. The implantation sites were mid and apical septum in 46 (90%) and 5 (10%) patients, respectively. Although RAO and LAO views suggested a septal location, 9 (17.6%) devices were found to be directing at the free wall in the left lateral view and required repositioning. One patient (2%) developed cardiac perforation due to contrast injection against the RV anterior wall before verification of sheath location by lateral view. Mean R-wave sensing and pacing threshold at implantation were 9.7 ± 4.0 mV and 0.61 ± 0.31 V/0.24 ms, respectively. After median follow-up of 218.7 days, the pacing threshold remained stable.

CONCLUSION

In this high-risk patient cohort, midseptal implantation of a leadless pacemaker as guided by RAO, LAO, and left lateral views was achieved in 90% of patients, with a low risk of complications.

Для цитирования: Hai Jo-Jo, Fang Jonathan, Tam Chor-Cheung, Wong Chun-Ka, Un Ka-Chun, Siu Chung-Wah, Lau Chu-Pak, Tse Hung-Fat. Safety and feasibility of a midseptal implantation technique of a leadless pacemaker.. Heart rhythm. 2018. DOI: 10.1016/j.hrthm.2018.12.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.