Noise insertion in CT for cocaine body packing: where is the limit of extensive dose reduction?
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Aissa Joel, Bölke Edwin, Sawicki Lino M, Appel Elisabeth, Thomas Christoph, Heusch Philipp, Sedlmair Martin, Krzymyk Karl, Kröpil Patric, Antoch Gerald, Boos Johannes
Аннотация
BACKGROUND: To evaluate the detection rate and image quality in CT-body-packer-screening at different radiation-dose levels and to determine a dose threshold that enables a reliable detection of incorporated body packs and incidental findings with a maximum of dose saving.
MATERIALS AND METHODS: We retrospectively included 27 individuals who underwent an abdominal CT with automated exposure control due to suspected body packing. CT images were reconstructed at different radiation-dose levels of 50%, 10, 5% and 1% using iterative reconstructions. All 135 CT reconstructions were evaluated by three independent readers. Reviewers determined the presence of foreign bodies and evaluated the image quality using a 5-point ranking scale. In addition, visualization of incidental findings was assessed.
RESULTS: A threshold of 5% (effective dose 0.11 ± 0.07 mSv) was necessary to correctly identify all 27 patients with suspected body packing. Extensive noise insertion to a dose level of 1% (0.02 ± 0.01 mSV) led to false-positive solid cocaine findings in three patients. Image quality was comparable between 100 and 50%. The threshold for correct identification of incidental findings was 10% of the initial dose (effective dose 0.21 ± 0.13 mSv).
CONCLUSIONS: Our results indicate that dose of abdominal CT for the detection of intracorporeal cocaine body packets can be markedly reduced to up to 5% of the initial dose while still providing sufficient image quality to detect ingested body packets. However, a minimum effective dose of 0.21 mSv (10% of initial dose) seems to be required to properly identify incidental findings.
MATERIALS AND METHODS: We retrospectively included 27 individuals who underwent an abdominal CT with automated exposure control due to suspected body packing. CT images were reconstructed at different radiation-dose levels of 50%, 10, 5% and 1% using iterative reconstructions. All 135 CT reconstructions were evaluated by three independent readers. Reviewers determined the presence of foreign bodies and evaluated the image quality using a 5-point ranking scale. In addition, visualization of incidental findings was assessed.
RESULTS: A threshold of 5% (effective dose 0.11 ± 0.07 mSv) was necessary to correctly identify all 27 patients with suspected body packing. Extensive noise insertion to a dose level of 1% (0.02 ± 0.01 mSV) led to false-positive solid cocaine findings in three patients. Image quality was comparable between 100 and 50%. The threshold for correct identification of incidental findings was 10% of the initial dose (effective dose 0.21 ± 0.13 mSv).
CONCLUSIONS: Our results indicate that dose of abdominal CT for the detection of intracorporeal cocaine body packets can be markedly reduced to up to 5% of the initial dose while still providing sufficient image quality to detect ingested body packets. However, a minimum effective dose of 0.21 mSv (10% of initial dose) seems to be required to properly identify incidental findings.