Journal of the American College of Cardiology · 1994
Transcatheter patent ductus arteriosus occlusion: application in the small child.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Nykanen D G, Hayes A M, Benson L N, Freedom R M
Аннотация
OBJECTIVES: This study intended to evaluate application of transcatheter occlusion of the patent ductus arteriosus in children < 10 kg body weight.
BACKGROUND: Transcatheter occlusion of the patent ductus arteriosus in the child weighing > 10 kg has been proved safe and effective.
METHODS: We reviewed 74 consecutive patients weighing < 10 kg (median 8.1 kg, range 3.98 to 10) and aged 4 to 30 months (median 13 months) who underwent patent ductus arteriosus occlusion between June 1986 and November 1992. A modification of the delivery system to facilitate application in small children is described.
RESULTS: A 12-mm device was implanted in 50 patients and a 17-mm device in 24. Three device embolizations occurred early in the experience, and one required removal because of hemolysis associated with a moderate residual shunt. One 17-mm device was removed at catheterization because of acute compromise to left pulmonary artery flow after implantation. Prevalence of residual shunting was 33% at 6 months, 20% at 12 to 18 months and 17% at 2-year follow-up and was not related to device, age, weight or size of the patent ductus. Altered flow to the left pulmonary artery was noted in seven patients (9.9%), with more significant compromise associated with the 17-mm device.
CONCLUSIONS: Transcatheter patent ductus arteriosus occlusion is feasible in the small child < 10 kg, particularly with the use of a modified delivery system. However, implantation of a 17-mm device in this patient population may impair flow to the left lung and should be reserved for the symptomatic child.
BACKGROUND: Transcatheter occlusion of the patent ductus arteriosus in the child weighing > 10 kg has been proved safe and effective.
METHODS: We reviewed 74 consecutive patients weighing < 10 kg (median 8.1 kg, range 3.98 to 10) and aged 4 to 30 months (median 13 months) who underwent patent ductus arteriosus occlusion between June 1986 and November 1992. A modification of the delivery system to facilitate application in small children is described.
RESULTS: A 12-mm device was implanted in 50 patients and a 17-mm device in 24. Three device embolizations occurred early in the experience, and one required removal because of hemolysis associated with a moderate residual shunt. One 17-mm device was removed at catheterization because of acute compromise to left pulmonary artery flow after implantation. Prevalence of residual shunting was 33% at 6 months, 20% at 12 to 18 months and 17% at 2-year follow-up and was not related to device, age, weight or size of the patent ductus. Altered flow to the left pulmonary artery was noted in seven patients (9.9%), with more significant compromise associated with the 17-mm device.
CONCLUSIONS: Transcatheter patent ductus arteriosus occlusion is feasible in the small child < 10 kg, particularly with the use of a modified delivery system. However, implantation of a 17-mm device in this patient population may impair flow to the left lung and should be reserved for the symptomatic child.