Clinica chimica acta; international journal of clinical chemistry · 2013
High throughput identification and quantification of 16 antipsychotics and 8 major metabolites in serum using ultra-high performance liquid chromatography-tandem mass spectrometry.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Patteet Lisbeth, Maudens Kristof E, Sabbe Bernard, Morrens Manuel, De Doncker Mireille, Neels Hugo
Аннотация
BACKGROUND: Therapeutic drug monitoring of antipsychotics is important for optimizing therapy, explaining adverse effects, non-response or poor compliance. We developed a UHPLC-MS/MS method for quantification of 16 commonly used and recently marketed antipsychotics and 8 metabolites in serum.
METHODS: After liquid-liquid extraction using methyl tert-butyl ether, analysis was performed on an Agilent Technologies 1290 Infinity LC system coupled with an Agilent Technologies 6460 Triple Quadrupole MS. Separation with a C18 column and gradient elution at 0.5 mL/min resulted in a 6-min run-time. Detection was performed in dynamic MRM, monitoring 3 ion transitions per compound. Isotope labeled internal standards were used for every compound, except for bromperidol and levosulpiride.
RESULTS: Mean recovery was 86.8%. Matrix effects were -18.4 to +9.1%. Accuracy ranged between 91.3 and 107.0% at low, medium and high concentrations and between 76.2 and 113.9% at LLOQ. Within-run precision was <15% (CV), except for asenapine and hydroxy-iloperidone. Between-run precision was aberrant only for 7-hydroxy-N-desalkylquetiapine, asenapine and reduced haloperidol. No interferences were found. No problems of instability were observed, even for olanzapine. The method was successfully applied on patient samples.
CONCLUSIONS: The liquid-liquid extraction and UHPLC-MS/MS technique allows robust target screening and quantification of 23 antipsychotics and metabolites.
METHODS: After liquid-liquid extraction using methyl tert-butyl ether, analysis was performed on an Agilent Technologies 1290 Infinity LC system coupled with an Agilent Technologies 6460 Triple Quadrupole MS. Separation with a C18 column and gradient elution at 0.5 mL/min resulted in a 6-min run-time. Detection was performed in dynamic MRM, monitoring 3 ion transitions per compound. Isotope labeled internal standards were used for every compound, except for bromperidol and levosulpiride.
RESULTS: Mean recovery was 86.8%. Matrix effects were -18.4 to +9.1%. Accuracy ranged between 91.3 and 107.0% at low, medium and high concentrations and between 76.2 and 113.9% at LLOQ. Within-run precision was <15% (CV), except for asenapine and hydroxy-iloperidone. Between-run precision was aberrant only for 7-hydroxy-N-desalkylquetiapine, asenapine and reduced haloperidol. No interferences were found. No problems of instability were observed, even for olanzapine. The method was successfully applied on patient samples.
CONCLUSIONS: The liquid-liquid extraction and UHPLC-MS/MS technique allows robust target screening and quantification of 23 antipsychotics and metabolites.