Randomised study comparing heart rate measurement in newly born infants using a monitor incorporating electrocardiogram and pulse oximeter versus pulse oximeter alone.

Murphy Madeleine C, De Angelis Laura, McCarthy Lisa K, O'Donnell Colm Patrick Finbarr

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

AIM: To determine whether IntelliVue (ECG plus Masimo pulse oximeter (PO)) measures heart rate (HR) in low-risk newborns more quickly than Nellcor PO (PO alone).

METHODS

Unmasked parallel group randomised (1:1) study.

RESULTS

We studied 100 infants, 47 randomised to IntelliVue, 53 to Nellcor. Time to first HR was shorter with IntelliVue ECG than Nellcor (median (IQR) 24 (19, 39) vs 48 (36, 69) s, p<0.001). There was no difference in time to display both HR and SpO2 (52 (47, 76) vs 48 (36, 69) s, p=0.507). IntelliVue PO displayed initial bradycardia more often than the Nellcor (55% vs 6%). Infants monitored with IntelliVue were handled more frequently and for longer.

CONCLUSIONS

IntelliVue ECG displayed HR more quickly than Nellcor PO. IntelliVue PO often displayed initial bradycardia. Infants monitored with IntelliVue were handled more often. Study of ECG in high-risk infants is warranted.

Для цитирования: Murphy Madeleine C, De Angelis Laura, McCarthy Lisa K, O'Donnell Colm Patrick Finbarr. Randomised study comparing heart rate measurement in newly born infants using a monitor incorporating electrocardiogram and pulse oximeter versus pulse oximeter alone.. Archives of disease in childhood. Fetal and neonatal edition. 2018. DOI: 10.1136/archdischild-2017-314366.