Comparison of device-specific adverse event profiles between Impella platforms.
Аннотация
BACKGROUND
The Impella (Abiomed) ventricular support system is a family of temporary mechanical circulatory support (MCS) devices used to treat patients with cardiogenic shock, acute cardiogenic decompensation, and for high-risk percutaneous or surgical revascularization. These devices include the percutaneously implanted 2.5/cardiac power (CP) and the surgically implanted 5.0/left direct (LD). Despite the beneficial effects and increased usage of these devices, data to assess adverse outcomes and guide clinician decision-making between the Impella CP and 5.0/LD are limited.
METHODS
This is a retrospective analysis of 91 consecutive patients who required at least 24 h of Impella support, from January 1, 2015 to December 31, 2019. Groups were stratified based on either initial Impella CP or 5.0/LD placement. Clinical outcomes and in-hospital complications were compared.
RESULTS
Impella CP was implanted in 66 patients (mean age: 61 ± 15 years, male 71.2%) and Impella 5.0/LD was implanted in 25 patients (mean age: 62 ± 9 years, male 84.0%). There was greater stability of device position (p = .033), less incidence of hemolysis (p < .001), and less frequent need for additional MCS (p = .001) in patients implanted with the Impella 5.0/LD compared with Impella CP in this study cohort. Patients with Impella 5.0/LD were more likely to survive from Impella and survive to discharge.
CONCLUSIONS
This study suggests that for patients who require temporary MCS for more than 24 h, the Impella 5.0/LD may have a more favorable device-specific adverse profile compared with the Impella CP.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.