DOI 10.1111/pace.13468

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2018

Subcutaneous implantable cardioverter defibrillator oversensing and shock delivery due to chest compressions during CPR.

Berkowitz Eric J, Pleimann Brent E, Rosenfeld Lynda E

Клинический случайСтатьяПеревод в очереди

Аннотация

We report the case of a patient who received both appropriate and inappropriate shocks from an entirely subcutaneous implantable defibrillator (S-ICD). The inappropriate shocks were due to oversensing of chest compressions in the setting of a profound postshock bradycardia. It is important to recognize that the therapy from the S-ICD can not only be withheld, as previously described, but can also be inappropriately delivered during chest compressions.

Для цитирования: Berkowitz Eric J, Pleimann Brent E, Rosenfeld Lynda E. Subcutaneous implantable cardioverter defibrillator oversensing and shock delivery due to chest compressions during CPR.. Pacing and clinical electrophysiology : PACE. 2018. DOI: 10.1111/pace.13468.