Treatment of thrombus formation associated with the MicroMed DeBakey VAD using recombinant tissue plasminogen activator.

Клиническое исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Rothenburger Markus, Wilhelm Markus J, Hammel Dieter, Schmidt Christoph, Tjan Tonny D T, Böcker Dirk, Scheld Hans H, Schmid Christof

Аннотация

BACKGROUND: The latest generation of left ventricular assist devices consists of nonpulsatile impeller pumps. In these small pumps, thrombus formation inside the device does not lead to thromboembolic end-organ dysfunction but may dramatically impair pump flow. We report on our experience with thrombus-related pump dysfunctions of the MicroMed DeBakey left ventricular assist device and its treatment.

METHODS: Eight of 22 patients with a MicroMed DeBakey VAD presented with a critically reduced pump flow. In 7 cases, an increased power demand indicative of progressive thrombus formation associated with the device was evident, whereas 1 case presented with thrombus formation within the inflow conduit associated with a very low power demand. Brief spontaneously resolving pump stops had been noted in 6 patients. All 8 patients were treated with 100 mg of recombinant tissue plasminogen activator (rt-PA), administered via an IV line.

RESULTS: Rt-PA lysis led to an increase of pump flow along with a reduction of power demand within a short time in all patients. No severe bleeding complications occurred. However, 4 patients experienced transient epistaxis. All patients could be discharged from intensive care immediately after discontinuation of thrombolytic therapy.

CONCLUSION: Rt-PA lysis is a very effective tool for thrombus-related pump dysfunction in patients with impeller pumps, which renders emergency surgical exchange unnecessary in most cases.