Anaesthesiology intensive therapy · 2020
An observational study comparing the performance of TOF-Cuff with TOF-Scan monitoring during anaesthetic induction in clinical routine.
Сравнительное исследованиеСтатьяНаблюдательное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Markle Andrew, Horn Katja, Welter JoEllen, Dullenkopf Alexander
Аннотация
INTRODUCTION: Neuromuscular monitoring by acceleromyography assesses the effects of non-depolarising neuromuscular blocking agents used during anaesthesia induction to optimise intubation conditions. A new type of neuromuscular monitor, TOF-Cuff, integrates electrode stimulation into a blood pressure monitoring cuff. Comparisons of this device with TOF-Scan, considered a clinical standard acceleromyography device, have not been published.
MATERIAL AND METHODS: This prospective, observational study was approved by the Ethics Committee East Switzerland (BASEC-nr. 2016-02044), and patients' consent was obtained before inclusion. The study's aim was to compare TOF-Cuff with TOF-Scan by measuring the duration from the administration of a neuromuscular blocking agent to a train-of-four (TOF) ratio of 0%. After anaesthesia induction, atracurium was administered (0.5 mg kg-1) and TOF ratios were recorded every 15 seconds using the two devices simultaneously. Patients were grouped according to body mass index (< or ≥ 30 kg m-2).
RESULTS: Twenty-five non-obese and twenty-five obese patients were included. In non-obese patients, bias was -3 s (± 21.2; limits of agreement -44.7 to 38.4; P = 0.702). In obese patients, bias was -20 s (± 35.0; limits of agreement -88.6 to 48.6; P = 0.0139). Large intra-individual differences of up to 60 seconds were detected even in non-obese patients.
CONCLUSIONS: A significant systematic difference in the time to reach a TOF ratio of 0% was found when using the two devices in obese patients. In non-obese and obese patients, there were large intra-individual and clinically relevant differences. The two devices cannot be used interchangeably.
MATERIAL AND METHODS: This prospective, observational study was approved by the Ethics Committee East Switzerland (BASEC-nr. 2016-02044), and patients' consent was obtained before inclusion. The study's aim was to compare TOF-Cuff with TOF-Scan by measuring the duration from the administration of a neuromuscular blocking agent to a train-of-four (TOF) ratio of 0%. After anaesthesia induction, atracurium was administered (0.5 mg kg-1) and TOF ratios were recorded every 15 seconds using the two devices simultaneously. Patients were grouped according to body mass index (< or ≥ 30 kg m-2).
RESULTS: Twenty-five non-obese and twenty-five obese patients were included. In non-obese patients, bias was -3 s (± 21.2; limits of agreement -44.7 to 38.4; P = 0.702). In obese patients, bias was -20 s (± 35.0; limits of agreement -88.6 to 48.6; P = 0.0139). Large intra-individual differences of up to 60 seconds were detected even in non-obese patients.
CONCLUSIONS: A significant systematic difference in the time to reach a TOF ratio of 0% was found when using the two devices in obese patients. In non-obese and obese patients, there were large intra-individual and clinically relevant differences. The two devices cannot be used interchangeably.
Предметные рубрики MeSH
AccelerometryAdultAgedAnesthesia, GeneralAtracuriumBlood Pressure DeterminationBody Mass IndexElectric StimulationFemaleHumansIntraoperative Neurophysiological MonitoringMaleMiddle AgedNeuromuscular Blocking AgentsNeuromuscular MonitoringNeuromuscular Nondepolarizing AgentsObesityPremedicationProspective Studies