MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ame2.12184

Медицинская библиотека · Animal models and experimental medicine · 2021

Safety and biodegradability of a synthetic dural sealant patch (Liqoseal) in a porcine cranial model.

Kinaci Ahmet, Bergmann Wilhelmina, van Thoor Sander, Redegeld Saskia, van der Zwan Albert, van Doormaal Tristan P C

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Liqoseal consists of a watertight layer of poly(ester)ether urethane and an adhesive layer containing polyethylene glycol-N-hydroxysuccinimide (PEG-NHS). It is designed to prevent cerebrospinal fluid (CSF) leakage after intradural surgery. This study assessed the safety and biodegradability of Liqoseal in a porcine craniotomy model.

METHODS

In 32 pigs a craniotomy plus durotomy was performed. In 15 pigs Liqoseal was implanted, in 11 control pigs no sealant was implanted and in 6 control pigs a control dural sealant (Duraseal or Tachosil) was implanted. The safety of Liqoseal was evaluated by clinical, MRI and histological assessment. The degradation of Liqoseal was histologically estimated.

RESULTS

Liqoseal, 2 mm thick before application, did not swell and significantly was at maximum mean thickness of 2.14 (±0.37) mm at one month. The foreign body reaction induced by Liqoseal, Duraseal and Tachosil were comparable. Liqoseal showed no adherence to the arachnoid layer and was completely resorbed between 6 and 12 months postoperatively. In one animal with Liqoseal, an epidural fluid collection containing CSF could not be excluded.

CONCLUSION

Liqoseal seems to be safe for intracranial use and is biodegradable. The safety and performance in humans needs to be further assessed in clinical trials.

Для цитирования: Kinaci Ahmet, Bergmann Wilhelmina, van Thoor Sander, Redegeld Saskia, van der Zwan Albert, van Doormaal Tristan P C. Safety and biodegradability of a synthetic dural sealant patch (Liqoseal) in a porcine cranial model.. Animal models and experimental medicine. 2021. DOI: 10.1002/ame2.12184.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.