BACKGROUND
Mpox (formerly monkeypox) is an emerging zoonotic disease of public health concern that presents as a rash mimicking other common viral exanthems. Unlike traditional testing algorithms relying on several assays, the BioFire FilmArray meningitis/encephalitis (ME) panel simultaneously detects common viruses causing rashes; however, Biofire ME is only licensed for testing on cerebral spinal fluid.
OBJECTIVES
This study evaluated use of the Biofire ME panel for detection and discrimination of herpes simplex virus types 1 and 2 (HSV-1 and HSV-2), varicella zoster virus (VZV), human herpesviruses type 6 (HHV-6), enteroviruses (EVs), and human paraechoviruses (HPeVs) from a dermal or mucocutaneous swabs collected in universal transport media (UTM).
STUDY
DESIGN
Results of the BioFire ME panel were compared against methods used during clinical testing. Ten-fold serial dilutions in UTM of cultured viruses were used to compare analytical sensitivity, and analytical specificity was assessed using panels of microorganisms in UTM. Clinical sensitivity and specificity were assessed using 20 positive specimens each for HHV-1, HHV-2, HHV-6, VZV, EVs, and HPeV, as well as 35 known negative specimens that included 15 mpox-positive specimens.
RESULTS
Biofire ME was as sensitive as comparator methods, and correctly discriminated all HSV-1, HSV-2, VZV, HHV-6, EVs, and HPeVs from mpox and mpox-mimickers. Cross-reaction between EV and rhinoviruses A, B, and C were noted in the specificity panel.
CONCLUSIONS
Swabs in UTM collected for mpox testing are suitable for use on the Biofire ME panel, allowing more streamlined diagnostic testing for viral exanthems in patients under investigation for mpox infection.