DOI 10.1097/pcc.0b013e3181bc5894
Pressure-created communication between lumens of a multilumen venous catheter.
Аннотация
OBJECTIVE
To describe the inadvertent creation of a connection between lumens of a multilumen catheter by use of an in-line intravenous hand pressure pump.
DESIGN
Case report.
SETTING
Operating theater.
PATIENT: All lumens of a 4F triple lumen catheter were verified patent by irrigation with saline before insertion into an internal jugular vein of an 18-mo-old child undergoing cardiac surgery. During surgery, the flushing of saline into one lumen of the catheter by means of an in-line hand pressure pump caused registration of >300 mm Hg pressure recorded from another lumen. The connections to these two lumens were then reversed. Repeated flushing also caused registration of pressure >300 mm Hg. On removal of the catheter at the end of surgery, flushing of one lumen caused fluid to emerge from the wrong exit hole as the appropriate exit hole was blocked with blood clot.
INTERVENTIONS: None.
MEASUREMENTS AND MAIN
RESULTS
Pressure measurement in intravenous catheter lumens. Greater than 300 mm Hg was recorded during flushing with an inline hand pressure pump.
CONCLUSIONS
High pressure created by use of an in-line intravenous hand pressure pump may cause creation of a communication between lumens of multilumen catheters.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.