INTRODUCTION

This study proposes guidelines for motor unit number index (MUNIX) recording and analysis.

METHODS

MUNIX was measured in control participants and in patients with amyotrophic lateral sclerosis. Changes in MUNIX values due to E1 electrode position, number of surface electromyography interference pattern (SIP) epochs, SIP epoch duration, force of contraction, and outlier data points were investigated.

RESULTS

MUNIX depends on optimized compound muscle action potential (CMAP) amplitude. Individual muscles showed variations when the number of epochs was low or when the SIP duration was short. Longer SIP duration allowed better recognition of artifacts. MUNIX results were affected by SIP values at all force levels but was more affected when SIP area was low.

DISCUSSION: We recommend changing the E1 electrode position to maximize CMAP amplitude. Twenty or more SIP signals of 500-ms duration should be recorded by using force levels ranging from slight to maximum. Traces should be reviewed to identify and exclude signals with tremor or solitary spikes. Muscle Nerve 58: 374-380, 2018.

Для цитирования: Nandedkar Sanjeev D, Barkhaus Paul E, Stålberg Erik V, Neuwirth Christoph, Weber Markus. Motor unit number index: Guidelines for recording signals and their analysis.. Muscle & nerve. 2018. DOI: 10.1002/mus.26099.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.