External implantable defibrillator as a bridge to reimplant after explant for infection: Experience from two centers.

Dell'Era Gabriele, Ziacchi Matteo, Prenna Eleonora, Diemberger Igor, Varalda Marco, Martignani Cristian, Guerra Federico, Biffi Mauro, Occhetta Eraldo

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

BACKGROUND

The management of patients explanted for implantable converter defibrillator (ICD) infections may be complex when anti-bradycardia pacing and tachyarrhythmia protection are needed. We aimed to test the efficacy and safety of a conventional ICD externally connected to a transvenous dual-coil lead as bridging therapy before the reimplantation.

METHODS AND

RESULTS

We enrolled seven patients explanted for ICD infections and needed prolonged antibiotic therapy in two high-volume hospitals in Italy and treated them with a passive-can external ICD for a mean of 13 (4-30) days before reimplant. One patient experienced an electrical storm, efficaciously recognized by the external ICD and treated with antitachycardia pacing and shocks. On-demand pacing was granted for all the patients. No device-related complications were reported.

CONCLUSIONS

An external ICD seems safe and efficacious as a bridge to reimplant in patients explanted for ICD infections.

Для цитирования: Dell'Era Gabriele, Ziacchi Matteo, Prenna Eleonora, Diemberger Igor, Varalda Marco, Martignani Cristian, Guerra Federico, Biffi Mauro, Occhetta Eraldo. External implantable defibrillator as a bridge to reimplant after explant for infection: Experience from two centers.. Pacing and clinical electrophysiology : PACE. 2018. DOI: 10.1111/pace.13320.