Utility of icteric index in clinical laboratories: more than a preanalytical indicator.

Mondejar Rufino, Mayor Reyes María, Melguizo Madrid Enrique, Cañavate Solano Consuelo, Pérez Ramos Santiago

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Total bilirubin tests are highly demanded in clinical laboratories. Since icteric index (I-index) has zero cost, we aimed to evaluate its clinical utility and cost-effectiveness to determine if total bilirubin is necessary to be tested. We took into account if haemolysis could interfere to icteric index determination.

MATERIAL AND

METHODS

Retrospectively we reviewed I-index results in two cohorts (43,372 and 8507 non-haemolysed and haemolysed samples, respectively). All determinations were done using Alinity c chemistry analysers (Abbott Diagnostics). Receiver operating characteristic (ROC) curve was used to determine the optimal index cut-off to discriminate between normal and abnormal bilirubin concentration (20.5 µmol/L).

RESULTS

The ROC curve analysis suggested 21.4 µmol/L as the optimal I-index cut-off but differences in sensitivity and specificity were detected between patient derivation. For rejecting purpose, 15.4 µmol/L and 17.1 µmol/L I-index thresholds were selected based on patient derivation (inpatients and emergency room; and primary care and outpatients, respectively) with 97% sensitivity and 0.25% false negative results. Sensitivity was much lower in haemolysed samples. We selected 34.2 µmol/L I-index as threshold to detect hyperbilirubinemia with 99.7% specificity and 0.26% false positive results, independent of haemolysis. With the icteric index cut-offs proposed, we would save 66% of total bilirubin requested and analyse total bilirubin in around 2% of samples without total bilirubin requested.

CONCLUSIONS

This study supports the use of I-index to avoid bilirubin determination and to identify patients with hyperbilirubinemia. This work considers that the economic and test savings could help to increase the efficiency in clinical laboratories.

Для цитирования: Mondejar Rufino, Mayor Reyes María, Melguizo Madrid Enrique, Cañavate Solano Consuelo, Pérez Ramos Santiago. Utility of icteric index in clinical laboratories: more than a preanalytical indicator.. Biochemia medica. 2021. DOI: 10.11613/bm.2021.020703.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.