Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology · 2021
Head-to-head validation of six immunoassays for SARS-CoV-2 in hospitalized patients.
СтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Zonneveld Rens, Jurriaans Suzanne, van Gool Tom, Hofstra Jorrit J, Hekker Thecla A M, Defoer Pien, Broekhuizen-van Haaften Patricia E, Wentink-Bonnema Ellen M, Boonkamp Lynn, Teunissen Charlotte E, Heijboer Annemieke C, Martens Frans, de Bree Godelieve, van Vugt Michele, van Houdt Robin, Amsterdam UMC COVID-19 Biobank
Аннотация
BACKGROUND: Detecting SARS-CoV-2 antibodies may help to diagnose COVID-19. Head-to-head validation of different types of immunoassays in well-characterized cohorts of hospitalized patients remains needed.
METHODS: We validated three chemiluminescence immunoassays (CLIAs) (Liaison, Elecsys, and Abbott) and one single molecule array assay (SIMOA) (Quanterix) for automated analyzers, one rapid immunoassay RIA (AllTest), and one ELISA (Wantai) in parallel in first samples from 126 PCR confirmed COVID-19 hospitalized patients and 158 pre-COVID-19 patients. Specificity of the AllTest was also tested in 106 patients with confirmed parasitic and dengue virus infections. Specificity of the Wantai assay was not tested due to limitations in sample volumes.
RESULTS: Overall sensitivity in first samples was 70.6 % for the Liaison, 71.4 % for the Elecsys, 75.4 % for the Abbott, 70.6 % for the Quanterix, 77.8 % for the AllTest, and 88.9 % for the Wantai assay, respectively. Sensitivity was between 77.4 % (Liaison) and 94.0 % (Wantai) after 10 dpso. No false positive results were observed for the Elecsys and Abbott assays. Specificity was 91.1 % for the Quanterix, 96.2 % for the Liaison, and 98.1 % for the AllTest assay, respectively.
CONCLUSION: We conclude that low sensitivity of all immunoassays limits their use early after onset of illness in diagnosing COVID-19 in hospitalized patients. After 10 dpso, the Wantai ELISA has a relatively high sensitivity, followed by the point-of-care AllTest RIA that compares favorably with automated analyzer immunoassays.
METHODS: We validated three chemiluminescence immunoassays (CLIAs) (Liaison, Elecsys, and Abbott) and one single molecule array assay (SIMOA) (Quanterix) for automated analyzers, one rapid immunoassay RIA (AllTest), and one ELISA (Wantai) in parallel in first samples from 126 PCR confirmed COVID-19 hospitalized patients and 158 pre-COVID-19 patients. Specificity of the AllTest was also tested in 106 patients with confirmed parasitic and dengue virus infections. Specificity of the Wantai assay was not tested due to limitations in sample volumes.
RESULTS: Overall sensitivity in first samples was 70.6 % for the Liaison, 71.4 % for the Elecsys, 75.4 % for the Abbott, 70.6 % for the Quanterix, 77.8 % for the AllTest, and 88.9 % for the Wantai assay, respectively. Sensitivity was between 77.4 % (Liaison) and 94.0 % (Wantai) after 10 dpso. No false positive results were observed for the Elecsys and Abbott assays. Specificity was 91.1 % for the Quanterix, 96.2 % for the Liaison, and 98.1 % for the AllTest assay, respectively.
CONCLUSION: We conclude that low sensitivity of all immunoassays limits their use early after onset of illness in diagnosing COVID-19 in hospitalized patients. After 10 dpso, the Wantai ELISA has a relatively high sensitivity, followed by the point-of-care AllTest RIA that compares favorably with automated analyzer immunoassays.