A multicentre evaluation of the safety of intracoronary optical coherence tomography.

Barlis Peter, Gonzalo Nieves, Di Mario Carlo, Prati Francesco, Buellesfeld Lutz, Rieber Johannes, Dalby Miles C, Ferrante Giuseppe, Cera Maria, Grube Eberhard, Serruys Patrick W, Regar Evelyn

Оценочное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

AIMS: Optical coherence tomography (OCT) is increasingly being applied to the coronary arteries. However, the risks associated with the imaging procedure are not yet well defined. The purpose of the present multicentre registry was to assess the acute complications associated with the clinical use of intra-coronary OCT in a large number of patients.

METHODS AND

RESULTS

Consecutive patients from six centres who had OCT examination were retrospectively included. All adverse events and complications, even if transient, were noted. Risks were categorised into: 1) self-limiting 2) major complications including major adverse cardiac events (MACE) and 3) mechanical device failure. A total of 468 patients underwent OCT examination for evaluation of: plaque (40.0%), percutaneous coronary intervention (28.2%) or follow-up stent tissue coverage (31.8%). OCT was performed using a non-occlusive flush technique in 45.3% with a mean contrast volume of 36.6+/-9.4ml. Transient chest pain and QRS widening/ST-depression/elevation were observed in 47.6% and 45.5% respectively. Major complications included five (1.1%) cases of ventricular fibrillation due to balloon occlusion and/or deep guide catheter intubation, 3 (0.6%) cases of air embolism and one case of vessel dissection (0.2%). There were no cases of coronary spasm or MACE during or within the 24 hour period following OCT examination.

CONCLUSIONS

OCT is a specialised technique with a relatively steep learning curve. Major complications are uncommon and can be minimised with careful procedural planning and having an awareness of the potential contributory risks, especially deep guide catheter intubation during contrast flushing. Upcoming developments will make OCT more practical and less procedurally demanding, also potentially conserving contrast volume considerably.

Для цитирования: Barlis Peter, Gonzalo Nieves, Di Mario Carlo, Prati Francesco, Buellesfeld Lutz, Rieber Johannes, Dalby Miles C, Ferrante Giuseppe, Cera Maria, Grube Eberhard, Serruys Patrick W, Regar Evelyn. A multicentre evaluation of the safety of intracoronary optical coherence tomography.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2009. DOI: 10.4244/eijv5i1a14.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.