MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/adc.2008.147900

Медицинская библиотека · Archives of disease in childhood. Fetal and neonatal edition · 2009

Neonatal percutaneous central venous lines: fit to burst.

Smirk C, Soosay Raj T, Smith A-L, Morris S

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To examine pressure changes in neonatal percutaneous central venous catheters under varying laboratory conditions and to quantify the risks of rupture in clinical practice.

DESIGN

We tested 27-gauge polyurethane Premicath and 24-gauge silicone ECC (both Vygon, Norristown, PA) catheters. Burst pressures were determined by applying a slowly ramped pressure to catheters that were occluded at the tip. Flow-pressure relationships were defined by increasing flow rates through patent catheters from 5 to 499 ml/h. Pressure changes during the manual flushing of catheters were determined for patent and occluded catheters and with different syringe sizes.

RESULTS

The mean burst pressure for polyurethane catheters (1730.8 kPa, 95% CI 1634.7 to 1826.8) was higher than for silicone catheters (275.6 kPa, 95% CI 240.4 to 310.8). Polyurethane catheters demonstrated an approximately fivefold greater margin of safety above manufacturer recommended operating pressures before burst compared to silicone catheters. Pressures remained at safe levels in both catheters over the range of flows generally used in neonatal practice. Hand-flushing of obstructed silicone catheters caused rupture in 5/6 silicone catheters tested, in comparison to 0/16 polyurethane catheters.

CONCLUSIONS

Polyurethane central venous catheters have a greater pressure tolerance than silicone catheters and are less likely to rupture under experimental conditions. Obstructed silicone catheters rupture easily when flushed. Catheters were not tested in human infants.

Для цитирования: Smirk C, Soosay Raj T, Smith A-L, Morris S. Neonatal percutaneous central venous lines: fit to burst.. Archives of disease in childhood. Fetal and neonatal edition. 2009. DOI: 10.1136/adc.2008.147900.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.