Failure to detect an unusual obstruction in a reinforced endotracheal tube with fiberoptic examination.

Paul Matthias, Dueck Michael, Kampe Sandra, Petzke Frank

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Obstruction of an endotracheal tube (ETT) is a potentially life-threatening event. We report an unusual obstruction of a reinforced ETT. The valve-like obstruction was caused by a partial detachment of the inner coating from the embedded spiral of the ETT. It led to an increase in inspiratory airway pressure, failure to detect end-expiratory CO(2), and generated a wheezing sound in forced expiration. Fiberoptic inspection, which is a recommended procedure for a suspected ETT-obstruction, failed to identify this detachment. Exchanging the defective ETT immediately resolved the critical clinical situation. The detachment was most likely caused by re-autoclavation of the ETT, which was a specified single-use product.

Для цитирования: Paul Matthias, Dueck Michael, Kampe Sandra, Petzke Frank. Failure to detect an unusual obstruction in a reinforced endotracheal tube with fiberoptic examination.. Anesthesia and analgesia. 2003. DOI: 10.1213/01.ane.0000076144.94487.bd.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.