DOI 10.1152/jappl.1978.44.4.619

Медицинская библиотека · Journal of applied physiology: respiratory, environmental and exercise physiology · 1978

Respiratory flow integrator with improved stability.

Deno N S, Barlett H L, Buskirk E R

СтатьяПоддержка службы здравоохранения СШАПеревод в очереди

Аннотация

A circuit was designed for integrating respiratory flow signals over time to produce volume signals. The circuit involves a simple integrator and three compensation subcircuits which control integrator drift caused by ambient temperature changes, flow signal base-line errors, and random signal errors. A functional unit built according to the circuit design requires initial adjustment, but no further adjustment before or during sustained use. Integration errors are typically less than 1%. The only flow signal requirements are that they be bidirectional and within the human physiological frequency range.

Для цитирования: Deno N S, Barlett H L, Buskirk E R. Respiratory flow integrator with improved stability.. Journal of applied physiology: respiratory, environmental and exercise physiology. 1978. DOI: 10.1152/jappl.1978.44.4.619.