Occurrence of significant long PR intervals in patients implanted for sinus node dysfunction and monitored with SafeR™: The PRECISE study.

Taieb Jérôme, Le Franc Pierre, Khattar Pierre, Moubarak Ghassan, Gorka Hervé, Godenir Jean-Philippe, Deutsch Philippe, Degand Bruno, Al Daoud Ayman, Elkaim Jean-Philippe, Elbaz Nathalie, Merlo Sandrine, Lellouche Nicolas

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Long PR intervals may increase cardiovascular complications, including atrial fibrillation. In pacemakers, the SafeR™ mode monitors PR intervals, switching from AAI to DDD when criteria for atrioventricular block are met.

AIMS: The PRECISE study evaluated the incidence and predictors of long PR intervals and their association with incident atrial fibrillation after 1 year in patients implanted for sinus node dysfunction and free from significant conduction disorders at baseline.

METHODS

This French, prospective, multicentre, observational trial enrolled patients implanted with a REPLY™ dual-chamber pacemaker. Pacemaker memory recorded long PR intervals (defined as first-degree atrioventricular block mode switches occurring after six consecutive PR/AR intervals≥350/450ms) and atrial fibrillation incidence (fallback mode switch>1minute/day). Predictors were identified from baseline variables (age, sex, AR and PR intervals, atrial rhythm disorder and medication) using logistic regression.

RESULTS

Of 291 patients with sinus node dysfunction enrolled, 214 were free from significant conduction disorders at baseline (mean age 79±8 years; 44% men; PR/AR intervals<350/450ms). After 1 year, long PR intervals had occurred in 116 patients (54%) and atrial fibrillation in 63 patients (30%). Amiodarone was the only independent predictor of long PR interval occurrence (odds ratio 2.50, 95% confidence interval 1.20-5.21; P=0.014). There was a strong trend towards an association between long PR interval and atrial fibrillation incidence (odds ratio 1.86, 95% confidence interval 0.97-3.61; P=0.051).

CONCLUSIONS

Half of the patients with pure sinus node dysfunction developed long PR intervals in the year following pacemaker implantation. Amiodarone was the only independent predictor of long PR intervals. There was a strong trend towards an association between long PR intervals and incident atrial fibrillation.

Для цитирования: Taieb Jérôme, Le Franc Pierre, Khattar Pierre, Moubarak Ghassan, Gorka Hervé, Godenir Jean-Philippe, Deutsch Philippe, Degand Bruno, Al Daoud Ayman, Elkaim Jean-Philippe, Elbaz Nathalie, Merlo Sandrine, Lellouche Nicolas. Occurrence of significant long PR intervals in patients implanted for sinus node dysfunction and monitored with SafeR™: The PRECISE study.. Archives of cardiovascular diseases. 2018. DOI: 10.1016/j.acvd.2018.09.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.