Sensitivity loss of Lif:Mg,Cu,P thermoluminescence dosemeters caused by oven annealing.
Аннотация
LiF:Mg,Cu,P is a very sensitive thermoluminescence material that can be used for personal dosimetry in radiology. But if this material is heated too much during annealing or reading it quickly loses its sensitivity. This study shows that an annealing oven in wide use can cause thermal damage to the dosemeters owing to inhomogeneous temperature distribution in the annealing tray. At annealing temperatures>240 degrees C, differences of only 1 degrees C lead to significant losses of sensitivity. Therefore, it is necessary to measure the distribution of temperature in the annealing tray for correct placement of the dosemeters in the tray.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.