Deactivation vs. asynchronous pacing - prospective evaluation of a protocol for rhythm management in patients with magnetic resonance conditional pacemakers undergoing adenosine stress cardiovascular magnetic resonance imaging.

Klein-Wiele Oliver, Garmer Marietta, Barbone Gianluca, Urbien Rhyan, Busch Martin, Kara Kaffer, Schäfer Harald, Schulte-Hermes Michael, Hailer Birgit, Grönemeyer Dietrich

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Cardiovascular Magnetic Resonance (CMR) imaging with adenosine stress is an important diagnostic tool in patients with known or suspected coronary artery disease (CAD). However, the method is not yet established for CAD patients with pacemakers (PM) in clinical practice. A possible reason is that no recommendations exist for PM setting (paused pacing or asynchronous mode) during adenosine stress. We elaborated a protocol for rhythm management in clinical routine for PM patients that considers heart rate changes under adenosine using a test infusion of adenosine in selected patients.

METHODS

47 consecutive patients (mean age 72.3 ± 10,0 years) with MR conditional PM and known or suspected CAD who underwent CMR in clinical routine were studied in this prospective observational study. PM indications were sinus node dysfunction (SND, n = 19; 40,4%), atrioventricular (AV) block (n = 26; 55.3%) and bradyarrhythmia in permanent atrial fibrillation (AF, n = 2; 4.3%). In patients with SND, normal AV-conduction and resting HR >45 bpm at the time of CMR and in AF the PM was deactivated for the scan. In intermittent AV-block a test infusion of adenosine was given prior to the scan. All patients with permanent higher degree sinuatrial or AV-block or deterioration of AV-conduction in the adenosine test were paced asynchronously during CMR, in patients with preserved AV-conduction under adenosine the pacemaker was deactivated. CMR protocol included cine imaging, adenosine stress perfusion and late gadolinium enhancement.

RESULTS

The adenosine test was able to differentiate between mandatory PM stimulation during CMR and safe deactivation of the device. In patients with permanent sinuatrial or AV-block (n = 11; 23.4%) or deterioration of AV conduction in the adenosine test (n = 5, 10.6%) asynchronous pacing above resting heart rate did not interfere with intrinsic rhythm, no competitive stimulation was seen during the scan. 10 of 15 (66,7%) patients with intermittent AV-block showed preserved AV-conduction under adenosine. As in SND and AF deactivation of the PM showed to be safe during CMR, no bradycardia was observed.

CONCLUSION

Our protocol for rhythm management during adenosine stress CMR showed to be feasible and safe and may be recommended for pacemaker patients undergoing routine CMR.

Для цитирования: Klein-Wiele Oliver, Garmer Marietta, Barbone Gianluca, Urbien Rhyan, Busch Martin, Kara Kaffer, Schäfer Harald, Schulte-Hermes Michael, Hailer Birgit, Grönemeyer Dietrich. Deactivation vs. asynchronous pacing - prospective evaluation of a protocol for rhythm management in patients with magnetic resonance conditional pacemakers undergoing adenosine stress cardiovascular magnetic resonance imaging.. BMC cardiovascular disorders. 2017. DOI: 10.1186/s12872-017-0579-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.