OBJECTIVE

To determine the sensitivity and specificity of the AdenoPlus test compared with real-time polymerase chain reaction (PCR) and to determine whether there was a reduction in antibiotic prescriptions with the use of AdenoPlus compared with the previous year.

PATIENTS AND

METHODS

A total of 125 patients with suspected infectious conjunctivitis were accrued from June 4, 2015, through September 27, 2015. Forty-six participants from the prospective cohort completed both AdenoPlus and PCR testing. Two hundred fifty age-matched individuals were in the retrospective cohort.

RESULTS

There was a significant reduction in the percentage of patients who received an antibiotic ophthalmic prescription in the prospective cohort vs the retrospective cohort (32% vs 45%; χ2P=.01). AdenoPlus test sensitivity was 50% (5 of 10) and specificity was 92% (33 of 36) compared with real-time PCR testing.

CONCLUSION

The AdenoPlus test has high specificity for diagnosing adenoviral conjunctivitis but lower sensitivity than has been previously published. These data suggest that negative AdenoPlus results should be confirmed by real-time PCR owing to the low overall sensitivity of AdenoPlus observed.

Для цитирования: Holtz Kelly K, Townsend Kalie R, Furst Joseph W, Myers Jane F, Binnicker Matthew J, Quigg Stephanie M, Maxson Julie A, Espy Mark J. An Assessment of the AdenoPlus Point-of-Care Test for Diagnosing Adenoviral Conjunctivitis and Its Effect on Antibiotic Stewardship.. Mayo Clinic proceedings. Innovations, quality & outcomes. 2017. DOI: 10.1016/j.mayocpiqo.2017.06.001.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.