A devices' game of thrones: cardiac resynchronization therapy vs. pacemaker.
Аннотация
AIMS: Oversensing can interfere with biventricular pacing. Cardiac Resynchronization Therapy (CRT) output inhibition due to automatic brady mode change from a sensing to a pacing mode of a previously implanted pacemaker as it reached battery capacity depleted indicator has not been previously published in the medical literature.
METHODS AND
RESULTS
We report the first case of CRT output inhibition in a pacemaker dependent patient due to electrical stimuli from a previously right-sided implanted pacemaker, after unaware reversion of OVO mode (O = no chambers paced; V = ventricular sensing; O = no response to sensing) to backup VVI (V = ventricular pacing; V = ventricular sensing; I = inhibitory response to sensing) when it reached the elective replacement interval.
CONCLUSION
This paper emphasizes the importance of knowing the distinct pacemaker brady mode behaviours after battery capacity depleted indicator has been reached, according to the pacemakers' manufacturer, including the possibility of automatic brady mode change from sensing to pacing mode. It also highlights the potential for severe bradycardia or asystole of this automatic brady mode change from a previously implanted pacemaker in pacemaker dependent patients submitted to CRT upgrade.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.