MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s10840-025-02049-0

Медицинская библиотека · Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2025

Preoperative skin antiseptics for the prevention of cardiac implantable electronic device infections: chlorhexidine-alcohol versus povidone-iodine-alcohol.

Da Costa Antoine, Ovache Lucas, Chellali Sarah, Guichard Jean Baptiste, Romeyer Cécile, Yvorel Cedric, Benali Karim

СтатьяНаблюдательное исследованиеСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Skin antisepsis plays a key role in the prevention of cardiac implantable electronic device (CIED) infections. Moreover, skin antisepsis by chlorhexidine-alcohol has not yet been fully evaluated.

OBJECTIVES

This single-center controlled study sought (1) to conduct a prospective observational analysis comparing two antiseptic skin preparations over two similar consecutive time periods, one conducted over a 1-year period using a povidone-iodine-alcohol solution (group I) and the other over the following year with a chlorhexidine-alcohol solution (group II) and (2) to determine real-life CIED infection predictive factors using the Prevention of Arrhythmia Device Infection Trial (PADIT) score.

METHODS AND

RESULTS

CIED implantations were performed in 1689 patients. A total of 17 patients (1.01%) developed a CIED infection. Long-term follow-up (16 ± 13 months) revealed no significant differences between groups: infections were observed in 9/844 patients (1.07%) in group I versus 8/845 patients (0.95%) in group II (p = 0.99). Univariate and multivariate logistic regression analyses were performed to identify infectious risk factors. Adjusted odds ratios (ORs) and hazard ratios (HRs) were calculated. Univariate analyses revealed that infection occurrence was positively correlated with acute heart failure (OR: 3.55; 95%CI [1.24-9.37]; p = 0.01 and HR: 3.37; 95% CI [1.28-8.87]; p = 0.014), renal failure (OR: 2.71; 95% CI [0.88-8.42]; p = 0.08 and HR: 2.99; 95% CI [0.97-9.17]; p = 0.05), use of temporary pacing before the implantation procedure (OR: 4.45; 95% CI [1.003-20.3]; p = 0.04 and HR: 4.35; 95% CI [0.99-19.05]; p = 0.05), PADIT score (OR: 1.3; 95% CI [1.064-1.54]; p = 0.009 and HR: 1.27; 95% CI [1.052-1.53]; p = 0.013), and PADIT score ≥ 5 (OR: 4.185; 95% CI [1.6-10.9]; p = 0.003 and HR: 3.91; 95% CI [1.51-10.149]; p = 0.005). No clear significant difference was found regarding the antiseptic skin preparation choice (OR: 0.88; 95% CI [0.34-2.3]; p = 0.8 and HR: 1.11; 95% CI [0.416-2.98]; p = 0.82). In multivariate analysis, the PADIT score ≥ 5 (OR: 3.18; 95% CI [1.125-8.99]; p = 0.003 and HR: 3.052; 95% CI [1.089-8.55]; p = 0.034) and use of temporary pacing before the implantation procedure (OR: 8.44; 95% CI [1.68-42.34]; p = 0.01 and HR: 8.49; 95% CI [1.732-41.69]; p = 0.008) were negatively correlated with the device infection risk.

CONCLUSION

This study found that chlorhexidine-alcohol and povidone-iodine-alcohol solutions displayed similar antiseptic effects regarding CIED infection prevention. Both use of temporary pacing before the implantation procedure and PADIT score ≥ 5 correlated with the device infection risk.

Для цитирования: Da Costa Antoine, Ovache Lucas, Chellali Sarah, Guichard Jean Baptiste, Romeyer Cécile, Yvorel Cedric, Benali Karim. Preoperative skin antiseptics for the prevention of cardiac implantable electronic device infections: chlorhexidine-alcohol versus povidone-iodine-alcohol.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2025. DOI: 10.1007/s10840-025-02049-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.