Long-term outcomes in leadless Micra transcatheter pacemakers with elevated thresholds at implantation: Results from the Micra Transcatheter Pacing System Global Clinical Trial.

Piccini Jonathan P, Stromberg Kurt, Jackson Kevin P, Laager Verla, Duray Gabor Z, El-Chami Mikhael, Ellis Christopher R, Hummel John, Jones D Randy, Kowal Robert C, Narasimhan Calambur, Omar Razali, Ritter Philippe, Roberts Paul R, Soejima Kyoko, Zhang Shu, Reynolds Dwight, Micra Transcatheter Pacing Study Group

Клиническое исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Device repositioning during Micra leadless pacemaker implantation may be required to achieve optimal pacing thresholds.

OBJECTIVE

The purpose of this study was to describe the natural history of acute elevated Micra vs traditional transvenous lead thresholds.

METHODS

Micra study VVI patients with threshold data (at 0.24 ms) at implant (n = 711) were compared with Capture study patients with de novo transvenous leads at 0.4 ms (n = 538). In both cohorts, high thresholds were defined as >1.0 V and very high as >1.5 V. Change in pacing threshold (0-6 months) with high (1.0 to ≤1.5 V) or very high (>1.5 V) thresholds were compared using the Wilcoxon signed-rank test.

RESULTS

Of the 711 Micra patients, 83 (11.7%) had an implant threshold of >1.0 V at 0.24 ms. Of the 538 Capture patients, 50 (9.3%) had an implant threshold of >1.0 V at 0.40 ms. There were no significant differences in patient characteristics between those with and without an implant threshold of >1.0 V, with the exception of left ventricular ejection fraction in the Capture cohort (high vs low thresholds, 53% vs 58%; P = .011). Patients with an implant threshold of >1.0 V decreased significantly (P < .001) in both cohorts. Micra patients with high and very high thresholds decreased significantly (P < .01) by 1 month, with 87% and 85% having 6-month thresholds lower than the implant value. However, when the capture threshold at implant was >2 V, only 18.2% had a threshold of ≤1 V at 6 months and 45.5% had a capture threshold of >2 V.

CONCLUSIONS

Pacing thresholds in most Micra patients with elevated thresholds decrease after implant. Micra device repositioning may not be necessary if the pacing threshold is ≤2 V.

Для цитирования: Piccini Jonathan P, Stromberg Kurt, Jackson Kevin P, Laager Verla, Duray Gabor Z, El-Chami Mikhael, Ellis Christopher R, Hummel John, Jones D Randy, Kowal Robert C, Narasimhan Calambur, Omar Razali, Ritter Philippe, Roberts Paul R, Soejima Kyoko, Zhang Shu, Reynolds Dwight, Micra Transcatheter Pacing Study Group. Long-term outcomes in leadless Micra transcatheter pacemakers with elevated thresholds at implantation: Results from the Micra Transcatheter Pacing System Global Clinical Trial.. Heart rhythm. 2017. DOI: 10.1016/j.hrthm.2017.01.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.