Changes in Peak Inspiratory Flow After Acute Bronchodilation: An Observational Study of Patients with Stable Chronic Obstructive Pulmonary Disease.
Аннотация
Introduction
Identifying factors influencing peak inspiratory flow (PIF) is essential for aerosol drug delivery in stable patients with chronic obstructive pulmonary disease. While a minimum PIF for dry powder inhalers (DPIs) is established, acute bronchodilator (BD) effects on PIF remain unknown. Materials and
Methods
An inspiratory flow meter (In-Check™ DIAL) was used to measure PIF in stable patients during a 24-week observational cross-sectional study. Additionally, bronchodilator responsiveness (BDR) was determined using the In-Check DIAL device and spirometry. Patients received four puffs of albuterol, and pre- and post-BD PIF, forced expiratory volume in one second (FEV1), and forced vital capacity were measured. Sixty-three patients completed acute BDR data collection from July 31, 2019, to November 9, 2021. Primary endpoints were pre- and post-BD spirometry and PIF. Statistical analyses included PIF correlations with FEV1. BD change was assessed according to inhaler resistance and sex (subgroup analysis).
Results
Median patient age was 64.8 years, 85.7% were non-Hispanic White, and 57.1% were female. The median increase in absolute PIF (In-Check DIAL) was 5.0 L/min, and the % PIF change was 8.9%. With albuterol, 57.1% experienced a PIF BD change >5.0%, whereas 49.2% experienced a change >10.0%. Similarly, 55.6% experienced an FEV1 BD change >5.0% and 28.6% had a >10.0% FEV1 BD change with albuterol. PIF was weakly correlated with FEV1 BD change (absolute; % PIF; r = 0.28 [p = 0.02]; r = 0.21 [p = 0.11]). Pre- and post-BD median PIF were 75.5 and 83.5 L/min for low-to-medium-resistance DPI and 45.0 and 52.0 L/min for high-resistance, respectively. The median increases in pre- and post-BD PIF were 9.0 L/min in males and 4.5 L/min in females. In contrast to when using the In-Check DIAL device, we observed no consistent bronchodilatory effects on PIF measured by spirometry.
Conclusions
Using the In-Check DIAL device, ∼50% of patients experienced >10% PIF increase after acute BD, potentially enhancing medication lung deposition. Further research is required to understand PIF's impact on medication delivery. ClinicalTrials.gov Identifier: NCT04168775.