Investigation of implantable signal transmission characteristics based on visible data of the human leg.

Gao Yue-Ming, Ye Yan-Ting, Lin Shi, Vasić Željka Lučev, Vai Mang-I, Du Min, Cifrek Mario, Pun Sio-Hang

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Signal transmission characteristics between implanted medical devices and external equipment has been a common key issue, as has the problem of supplying energy to the devices. It can be used to enable signal transmission from implanted devices that the human body's conductive properties. Using signal transmission by galvanic coupling is one of the most effective signal transmission methods.

METHODS

The signal transmission characteristics by galvanic coupling of implantable devices using a frequency range of 10 kHz to 1 MHz was analyzed in this article. A finite element (FEM) model and a phantom model established by visible human leg data were used to investigate the signal transmission characteristics of implant-to-surface, with implantable receiver electrodes at different locations.

RESULTS

The results showed that the FEM model and the phantom model had similar implantable signal transmission characteristics, with an increase of frequency, signal attenuation basically remained unchanged. The gain in signal attenuation in the fixed attenuation values fluctuated no more than 5 dB and signal attenuation values rose as the channel length increased.

CONCLUSIONS

Our results of signal transmission characteristics of surface-to-implant will provide a theoretical basis for implantable transceiver design, and for realization of a recharging method for implanted medical devices.

Для цитирования: Gao Yue-Ming, Ye Yan-Ting, Lin Shi, Vasić Željka Lučev, Vai Mang-I, Du Min, Cifrek Mario, Pun Sio-Hang. Investigation of implantable signal transmission characteristics based on visible data of the human leg.. Biomedical engineering online. 2017. DOI: 10.1186/s12938-017-0379-y.